| CompanyProductsOrderingResearch ApplicationsNews, Resources & Events |
Catheter UseVia a catheter, ALZET pumps can deliver substances into the venous or arterial circulation, into the brain, or into any organ, lumen, or solid tissue (List of major routes of administration). Attachment to a catheter does not alter the delivery rate of the pump. A variety of catheter materials have been used successfully with the pumps. To operate the pump with a catheter, perform the following steps.
100 µl and 200 µl pumps before and after the attachment of catheters
Technical Information for catheter applicationsImpact on Release RateThe correct use of a catheter with an ALZET pump does not change the release rate of the pump. However, there are situations involving the use of a catheter which can affect the rate at which drug solution is released from the end of the catheter:
Choice of tubingTypes of tubing which have been used with the pump include: Polyethylene tubing Vinyl tubing Polyurethane tubing
of a fine gauge (such as 32-gauge) Silastic®
tubing Tubing SizesThe following sizes of polyethylene (PE) and vinyl (V) catheter tubing fit well onto the flow moderators of all models of ALZET pumps:
|
| Length of Tubing |
Tubing Type |
Volume Contained |
| 1 cm |
PE-60 |
4.566 μl |
| 1 cm |
PE-50 |
2.679 μl |
| 1 cm |
V3/A |
3.739 μl |
| 1 cm |
PE-10 |
0.615 μl |
Following are tubing lengths required to hold the entire reservoir volume* of an ALZET pump:
| Pump Volume* |
PE-60 Length |
PE-50 Length |
V3/A Length |
| 100 µl |
22 cm |
37.3 cm |
26.74 cm |
| 200 µl |
43.8 cm |
74.7 cm |
53.49 cm |
| 2000 µl (2 ml) |
438 cm |
746.5 cm |
534.9 cm |
*Based on nominal fill volumes. The exact volume needed depends on the mean fill volume of the specific lot of pumps being used as printed on the instruction sheet supplied in each box of pumps.
These conversions are calculated using the formula for a volume of a cylinder:
πR2 x length = tubing volume
In this equation, R is the radius of a cross-section of the tubing, or half of the inner diameter given in section 3 above. The length is equal to the length of tubing being considered. Be sure to use the same units of measure throughout, and to convert the volume to μl to relate the final volume to the flow rate of the pumps, which is in μl/hr.
Pumps can be used with or without catheters. For catheter applications, the flow moderators supplied with all ALZET pumps will accept any tubing that will grip a No. 21 gauge needle (e.g. PE-60 tubing). If a smaller catheter, such as PE-10 (which grips a 30 gauge needle), is desired, there are two ways of achieving the necessary step-down in catheter size.
Polyethylene tubing can easily be drawn out over a heat source to a very small size. Hold the PE tubing about 6" above a low flame, and rotate the tubing while moving it back and forth across the flame, pulling steadily. After the tubing has been drawn out, attach a water-filled syringe to the unflamed end, using a 21 gauge needle. Apply pressure on the plunger, and using scissors cut off the attenuated end in successive pieces until water flows out. This is the smallest diameter you can reasonably obtain with this procedure.
We have successfully obtained open tips no larger than 4x the diameter of a red blood cell by this means, while the base end of the catheter retained its original geometry to fit on a 21 gauge needle.
Another method for using very
small bore catheters with ALZET pumps is to glue PE-10 tubing inside
PE-60. Be sure to test for leakage at this junction before implanting.
(Sources for purchasing PE-10 tubing)
| Tubing Type |
Inner Diameter |
Outer Diameter |
| PE-10 |
0.28 mm |
0.64 mm |
Some agents are very expensive or available only in volumes less than 100 µl, which is the reservoir capacity of the smallest ALZET pump. The pump reservoir cannot be filled to less than capacity, since doing so would result in a reservoir filled partially with solution and partially with air, causing discontinuous delivery of the solution. The ALZET pump is easily adapted for delivery of smaller volumes by attaching a small “external reservoir” made of catheter tubing.
For example, to deliver 25 µl of solution, one would fill a short piece of catheter tubing just long enough to hold 25 µl of solution. This tubing would then be attached to an ALZET pump completely filled with a control solution (e.g., saline). In this example, 5.5 cm of PE-60 tubing would be sufficient. Refer to Catheter Volumes above to calculate the length of tubing needed. If the required tubing will be too long and unwieldy for implantation, refer to Preparing a Lynch Coil to make it more space efficient.
If mixing between the control and test solutions is a concern, one of the following may be done:
The Lynch coil method can be used for several purposes:
Polyethylene tubing should be used because it is thermoformable and can be formed into a permanent coil. The length of tubing depends upon the size and duration of pump being used, and the duration of infusion desired from the solution in the coil. The pump itself is filled with saline, which is separated from the drug solution by a drop of mineral oil or other nonmiscible compound or a bubble of air placed in the tubing at the end between the coil and the pump.
Lynch coils attached to ALZET Osmotic Pumps

|
|
||
![]()
©2009 DURECT Corporation – All Rights Reserved. |