Sterility: Solutions and PumpsBecause the ALZET pump is designed to contain solutions for as long as six weeks (depending on the model) at 37°C, any microorganisms present in the loading solution may grow, metabolize, and alter the composition of the test solution. This risk is higher if the solution contains materials such as proteins that are potential substrates for microorganisms. For this reason, all solutions loaded into the pumps should be sterile.
Maintaining sterility of solutionsCustomers have reported instances in which the biological activity of an agent delivered by the pump seemed to fade or disappear as the experiment progressed. While this observation could be a real and important aspect of the pharmacology of the substance in some cases, in other cases it suggests an inoculation of the test solution with microorganisms.
Maintaining pump sterilityALZET pumps are supplied sterile, having been exposed to a sterilizing dose of 60Co. During use, however, exterior contamination from handling may occur. Therefore, we recommend using sterile technique whenever handling the pumps. After loading, the surface of the pump can be cleaned by wiping it with isopropyl alcohol (70% in water). If this optional step is taken, avoid getting alcohol into the delivery port of the pump. The pump should not be soaked in alcohol.
|
|
||||||||||||||||||